How to replace
Removal
1
• Remove the following covers:
Lamp cover
A
and back cover
B
, (section 4•2 steps 1-2).
2
• Remove the hardware unit, (section 4•5, steps 2-6).
3
• Disconnect connector J20, (green)
or
J21, (black) from the control board.
Note:
earlier model scanners use J20, (green)- see the note in step 11.
4
• Loosen the three 3 mm set screws
A
on the sensor module, securing
the three RGB fibre optics cables, and carefully remove the cables.
Note:
mark the cables as necessary to ensure correct refitment.
5
• Loosen the 3 mm allen screw
B
securing the PMT module to its tray
and slide the module out.
Assembly
6
• Fit the three fibre optics cables (as marked in step 4) into the sensor
module and tighten the 3 mm set screws
A
.
Note
: Do not bend the fibre optics cable excessively when fitting.
7
• Connect connector J20, (green)
or
J21, (black) to the control board.
Note:
If J20, (green) is used, it is possible to fit it both ways. If fitted in-
correctly, it will cause the amplifiers in the PMT module to blow. Ensure
that the locking tabs on the connector fit into the locks on the control
board, (that is, facing to the
right hand side
viewed from the
back
of the
scanner).
8
• Carry out a PMT amplifier gain adjustment, (section 5•6).
9
• Fit the new PMT module by positioning the board onto its tray and
carefully pushing in.
10
• Tighten the 3 mm allen screw
B
.
11
• Partially refit the hardware unit, (section 4•5 steps 7-10).
Note:
if J20 (green) is used on the control board, ensure that is con-
nected correctly, (see the caution on the previous page, 4•11a).
12
• Carry out the adjustment and test operations specified below, (except
PMT amplifier gain as this was carried out in step 8).
13
• Carefully push the hardware unit fully in, (connector J1 on the driver
board mates with the backplane) and fit the two 2.5 mm countersunk
allen screws
A
on each side of the unit.
14
• If no further replacement operations are required, refit the covers
removed i n step 1 in accordance with section 4•2.
Adjustments required
PMT amplifier gain adjustment, (section 5•6).
White calibration in transmission and reflection (section 5•5).
Register error (section 5•12) and stripes in shadow (section 5•13).
Service Manual • ScanMate 4000/5000 Edition 1
•
January 1996
PMT Module
4•11b
Summary of Contents for ScanMate 4000
Page 1: ...ScanMate 5000 ScanMate 4000 ...
Page 6: ......
Page 16: ......
Page 18: ......
Page 20: ......
Page 22: ......
Page 24: ......
Page 28: ......
Page 40: ......
Page 44: ......
Page 48: ......
Page 52: ......
Page 56: ......
Page 60: ......
Page 64: ......
Page 68: ......
Page 72: ......
Page 76: ......
Page 80: ......
Page 84: ......
Page 88: ......
Page 92: ......
Page 96: ......
Page 100: ......
Page 102: ......
Page 106: ......
Page 110: ......
Page 114: ......
Page 118: ......
Page 122: ......
Page 126: ......
Page 130: ......
Page 134: ......
Page 138: ......
Page 142: ......
Page 146: ......
Page 150: ......
Page 154: ......
Page 156: ......
Page 162: ......
Page 164: ......
Page 170: ......
Page 171: ...ScanMate 4000 Type 250 Circuit Connection Diagram ...
Page 172: ...ScanMate 5000 Type 210 Circuit Connection Diagram ...
Page 174: ......