NSG 438
22
4.2.6.
Measurement adapters
The measurement adapter type MD 101 as per IEC / EN 61000-4-2 (2001)
serves to verify pulse amplitudes and pulse shapes. It is designed for
mounting in the side wall of a Faraday cage in which an oscilloscope has
been installed. This measurement adapter has the flat impedance curve to
well over 1GHz that is necessary for the purpose.
Use of this adapter is only worthwhile in conjunction with a test rig that is
laid out in strict conformity with the relevant standard (see section
“Verification of pulse data”).
MD 102 (“Pommerenke” target) is a more advanced coaxial measuring
target with flat response characteristics up to the multi-GHz range. It may
be used instead of MD 101.
MD 102