NSG 438
33
6.3.10. Threshold
This function permits differing sensitivity levels to be set for the arcing
detector whereby a differentiation can be made between stray discharges
and a true discharge onto the EUT.
Normal:
Arcing is detected and is indicated by the
kV
symbol on the
display blinking provided 20% (or more) of the charge voltage
is dissipated.
Low:
Arcing is detected and is indicated by the
kV
symbol on the
display blinking provided 10% (or more) of the charge voltage
is dissipated.
High:
Arcing is detected and is indicated by the
kV
symbol on the
display blinking provided 30% (or more) of the charge voltage
is dissipated.
Off
:
This position is made for EUTs with non-conductive surfaces
(housings).
This means in contact discharge mode, the arcing detection is disregarded
and the counter counts up/down as per settings in the repetition or counter
menu respectively.
In air discharges, this feature is available only in single mode. Pushing the
trigger once, the HV is activated and it can approach to the EUT until an
arcing happens or the EUT being touched. Then it has to push the trigger
a second time resulting in HV off and increasing/decreasing the counts,
depending on counter setting.
The “Off” operating function is included in NSG 438 products from serial
number 521 (March 2006) upwards.
Threshold
Off
Discharge detection