6.3.11. ISO-Selftest
Just starting the ISO-Selftest gives a quick response of the proper
operation of the ESD generator. The screen reflects all required voltage
levels as well as the tolerances in table form, based on the ISO 10605
standard.
During the calibration procedure the HV module is strained up to 30kV. If
somehow the maximum voltage could not be reached or hold, the NSG
438 come out correctly with an HV error message. The calibration
procedure is diagnosing, to which voltage level the HV module will work
properly. This voltage value will be shown on screen while it will not be
possible to set a higher voltage value.
This extended operating function is included in NSG 438 products from
serial number 521 (March 2006) upwards
.
Settings
ISO-Selftest
ISO-Selftest
kV 2 4 6 8 15 25
± 500V 10%
+
-