Self-calibration - fault correction
If the values of the self-calibration
exceed a
certain limit, the procedure is stopped for
reasons of safety.
For this fault, a pollution of the contacts at the
bottom of the measuring chamber may be
responsible.
In this case slightly sand the
stainless steel contacts (marked in blue on
the picture) using 400 grade abrasive paper.
Afterwards the contacts have to be cleaned
with alcohol.
If after that procedure the self-calibration is still not successful,
please contact Schaller GmbH.
Note
:
For the self-calibration the
measuring chamber has to be empty and cleaned!