Most common reasons for incorrect measurements

 

 

Product temperature out of application range

 

Material 

below  0°C 

resp. 

above  +50°C 

may  cause  faulty 

measurements.  

 

Measuring material is frozen or mixed with snow  

In these cases the accuracy is decreasing heavily. 

 

Wrong calibration curve 

Before  starting  a  measurement  please  double-check  if  you 
have chosen the right calibration curve (material).  

 

Wet or mouldy material 

In this case the accuracy may decrease. 

 

Moisture measurement of barks 

In  case  of  uneven  shape  of  barks  in  comparison  of  wood 
chips, the accuracy may decrease. 

 

Wrong filling quantity 

Ensure  that  the  measuring  box  is  filled  with  material 
completely; there must not be less neither more material in it. 

 

Wrong filling procedure 

Place  the  measuring  box  on  the  floor  and  pour  the  material 
into it. Use delivered bucket for that action at every time. 

Device maintenance instructions 

To  provide  a  long  life  of  your  device  please  do  not  expose  it  to 
strong mechanical loads, extreme heat, rain and heavy dust.  
The  weighing  plate  (grey  plastic  plate  at  the  bottom  of  the  BMA) 
has  to  be  free  from  wood  chips and  dirt.    In 
case  of  pollution  please  vacuum-clean  the 
weighing plate.  

Mind  the  golden  contacts,  they  must  not  be 
touched! 

At  the  bottom  of  BMA  housing  is  a  cover 
plate for cleaning screwed. You can remove that and vacuum-clean 
inside.Clean the measuring box using a 

dry cloth or a soft brush

Any  kind  of  wet  cleaning  and  any  use  of  cleaning  powder  may 
damage the device. The 

device  and the  measuring  box are not 

rainproof

, keep them in dry areas!  

Summary of Contents for humimeter BMA

Page 1: ...User manual Moisture meter for determining water content of biomass humimeter BMA version 1 8 Schaller GmbH 2014...

Page 2: ...2 version 1 8 Design of the device display start push button ON OFF button LED status power supply interface measuring box drawer temperature bulk density option water content atro weight m option mat...

Page 3: ...ply unit at the humimeter BMA and tighten it 4 Plug in the power supply unit into a Schuko socket 230VAC 5 After plugging in the humimeter BMA switches on automatically If not switch on the instrument...

Page 4: ...Select the right product calibration curve using the arrow keys 6 Press the Start push button The measurement will start automatically During the measurement the LED is blinking in blue and the displa...

Page 5: ...by performing several measurements 3 with the same sample and calculating the average of them If the individual measuring values are stored on the BMA the device calculates the average automatically...

Page 6: ...ht calibration curve If you are not sure about the right calibration curve we recommend carrying out a comparison measurement by kiln drying EN 14774 At page 12 and 13 you find pictures of different s...

Page 7: ...l logs Clear logs Send logs Manual logs Clear logs Options Status Options Date Time Datalog Time Language Unlock C F o Userlevel BL On time Auto Off Time Calibrate Materialcalib Adjust Password Reset...

Page 8: ...transfer on the humimeter Press the key until you reach the menu see image on the right Then choose Send Logs and confirm by pressing the key Now choose Manual Logs and confirm with again All saved lo...

Page 9: ...he selected logs are printed out now To save paper please think of clearing the data storage regularly If the printing has been started successfully the printer begins to print out the logs immediatel...

Page 10: ...ber and confirm by pressing Online print 2 times Options o Online Print For choosing that menu item you have to unlock your BMA first If you have activated that function your BMA will print the curren...

Page 11: ...version 1 8 page 11 Definition of important units for wood chips accounting...

Page 12: ...measuring chamber may be responsible In this case slightly sand the stainless steel contacts marked in blue on the picture using 400 grade abrasive paper Afterwards the contacts have to be cleaned wit...

Page 13: ...version 1 8 page 13 Coarse ch Industr WC...

Page 14: ...within a short time Considering a sampling error due to the considerably smaller sample quantity as well as the content of volatile matters resin etc that are not water the kiln drying method will pra...

Page 15: ...tic Power supply 100 240 VAC 1 A 50 60 Hz Plug Schuko plug as per CEE 7 7 Display 128x64 matrix display lighted Dimensions BxTxH 432 x 282 x 862 mm Weight approx 26 kg with measuring box Degree of pro...

Page 16: ...letely there must not be less neither more material in it Wrong filling procedure Place the measuring box on the floor and pour the material into it Use delivered bucket for that action at every time...

Reviews: