Checking the calibration 

 

We recommend to check the calibration of your humimeter PM3 every 4 weeks, using the 
proof plate delivered with the device. For this check the device has to show a temperature 
between  18,0 and 24,0°C. 

 

Switch  on  the  instrument  and  select  the  calibration  curve 
„reference“ using the arrow keys. Hold the humimeter PM3 
in  one  hand  and  press  it  onto  the  grey  proof  plate 
(compacting  pressure  of  1,0kg).  Make  sure  that  you  hold 
the device horizontally and the steel clamps are positioned 
at least 1 cm from the edge of the proof plate. 

 

The  shown  measuring  value  should  range  between 

18,5 

and

 20,5

. If the shown value is out of this range, you have 

to carry out a calibration (see instructions below).

  

  

Calibration 

By  a  zero  point  adjustment  you  can  correct  a  too  high 
deviation at the calibration check. 

For that please proceed as follows: 

Press the left key twice to reach the menu level. Navigate 
to the menu item „Options“ using the arrow  keys. Confirm 
by pressing the   button.  

Select the menu item „Calibrate“ using the arrow keys and 
confirm by pressing  .  

This  menu  item  has  to  be  unlocked.  For  that  you  have  to 
enter  the  4-digit  serial  number  of  your  humimeter  PM3. 
You can find the serial number on the top right edge of the 
display a short time after switching on the instrument, or on 
the sticker in the battery slot. 

Press  the  third  button  as  long  as  the  black  bar  stands  on 
the first digit of the serial number. After a waiting time of a 
few seconds the device accepts the entered digit.  Now the 
other 3 digits have to be entered. When the 4-digit number 
has  been  entered  correctly  (see  picture),  press  the 

 

button to confirm. 

Hold  the  humimeter  PM3  in  one  hand  up  in  the  air  and 
make  sure  that  nothing  stands  behind  the  steel  clamps 
(distance min. 200 mm).  Now press the second button ( ) 
with one finger.   

The  calibration  procedure  will  take  a  few  seconds.  During 
that  time,  the  instrument  has  to  be  held  up  in  the  air  and 
the measuring field (steel clamps) must NOT be touched.  

The  device  is  ready  when  the  shown  picture  on  the  right 
disappears automatically.

 

Summary of Contents for humimeter PM3

Page 1: ...User manual humimeter PM3 paper moisture meter version 1 2_en Schaller GmbH 2012 ...

Page 2: ...nning rolls the PM4 contact free temperature measurement is necessary 4 Now the display shows the water content The temperature is shown on the left 5 To save the results in the save menu press the button The storage was successful when the number in front of the symbol increased To reach the store menu please press the left button 6 The menu level Options Log time offers the possibility to activa...

Page 3: ...es that offer realistic results and write down the measuring results of the different calibration curves 2 Now please note the effective water content determined by your online measurement system or carry out a reference measurement according to EN ISO 287 3 Compare the determined reference water content with the measuring results of the different calibration curves Use the calibration curve with ...

Page 4: ... Copy paper calendered 1100 kg m 1200 paper LWC calendered 1200 kg m 1300 paper Flysheet paper 1300 kg m 1400 paper Flysheet paper 1400 kg m 1500 paper Flysheet paper high density 1500 kg m 1600 paper Flysheet paper very high density 1600 kg m Empty 1700 For creation of customer calibration curve Empty 975 For creation of customer calibration curve Empty 550 For creation of customer calibration cu...

Page 5: ...with a dry cloth For cleaning the steel clamps you can also use alcohol Please do not deform the steel clamps If the device is not used for a longer period 2 months or when the batteries are empty they should be removed to prevent a leakage of the battery acid name of calibration curve rolling menu button protective rubber housing removable ON OFF button measuring value water content measuring val...

Page 6: ...tch upper Switch lower Save Hold Watch saved data Add supplier s data menu Enter Switch upper Switch lower Back Enter numbers Enter letters Next or right Left Yes No Shift OK Menu level overview Data storage Manual logs Auto logs Clear logs Print logs Last logs All logs Clear logs Send logs Manual logs Auto logs Clear logs Options Options Date Time Log Time Language Unlock C F Calibrate o Userleve...

Page 7: ...menu item Options using the arrow keys Confirm by pressing the button Select the menu item Calibrate using the arrow keys and confirm by pressing This menu item has to be unlocked For that you have to enter the 4 digit serial number of your humimeter PM3 You can find the serial number on the top right edge of the display a short time after switching on the instrument or on the sticker in the batte...

Page 8: ...after switching on the batteries have to be changed IMMEDIATELY Also remove the batteries if you do not use your humimeter device for a longer period For eventual resulting damages we cannot provide any warranty Exemption from liability For miss readings and wrong measurements and of this resulting damage we refuse any liability This is a device for quick determination of moisture The moisture dep...

Page 9: ...e data transfer on the humimeter Press the key until you reach the menu see image on the right Then select Send Logs and confirm by pressing the key Now select Manual Logs and confirm by pressing again All saved logs will be sent to your PC Starting the data transfer on your PC Press the button remote control in the LogMemorizer software A drop down menu with several options opens see image below ...

Page 10: ...M3 Then switch on the device Not till then the other side of the cable has to be plugged in at the printer Switch on the printer by pressing Now the green LED is blinking If it does not blink please change the batteries and try again Press the button at your humimeter until you reach the menu see image on the right Select Print Logs and confirm by pressing Now you can select if you want to print t...

Page 11: ...x 10 000 values Menu languages German English French Italian Spanish Russian Power supply 4 pcs 1 5 Volt AA Alkaline batteries for approx 900 measurements Switch off time after approx 4 minutes Power consumption 60 mA with display lighting Display 128 x 64 matrix display with LED backlighting Dimensions 182 x 75 x 54 mm with rubber housing Weight 330g incl batteries and rubber housing Protection c...

Page 12: ...of application range The application range is between 0 and 60 C Discrepancy in temperature between device and material Please ensure that the device and the material under test are being stored at nearly the same temperature before measuring A high temperature difference has a negative effect on the stability of the measurement results Measurement at the front side A measurement at the front side...

Reviews: