Page 41
3. Use the arrow keys to select "Reference" under product type (see
).
4. Take hold of the device handles on both sides and
press it onto the top of the grey test plate inside
the case with a pressure of approx. 4.0 kg (figure
»
The moisture content reading shown must be be-
tween 14.5 and 15.5. (The moisture reading will be
displayed in black) (figure
»
If the moisture value is outside this range, in which
case it will be displayed in grey (figure
moisture meter has to be calibrated (see
11.5 Calibrating the moisture meter
11.5.1 Automatic calibration
11.5.2 Starting the calibration manually
1. Press
twice or hold for 2 seconds.
2. Select
Options
. To do so, press
or
and con-
firm by pressing
.
3. Select
Reinitialize
. To do so, press
or
and
confirm by pressing
.
»
The display will then show the message
Adjust?
(figure
).
4. Lift the device up into the air with both hands,
holding onto both handles. When doing so, there
must be a minimum of 0.5 metres of empty space
behind the sensor bars (figure
).
5. Confirm by pressing
.