User manual humimeter PMZ
Page 16
4. To ensure a correct measurement, the above mentioned pulp height must be
maintained to ensure a minimum contact pressure on
the electrodes.
»
The measured value is immediately visible on the
instruments display (figure
13
).
»
Now you can save the displayed measured value on
the device (see
"5.4 Saving individual readings"
or
"5.5 Saving several readings (measurement series) at the same time")
.
ATTENTION
Injury hazard
Injury hazard due to measuring electrodes.
►
Keep the measuring electrodes away from your body during all activities.
►
Keep the measuring electrodes away from other people' bodies during all
activities.
i
Information - Measuring accuracy
This rapid and non-destructive measuring procedure allows you to quickly take
moisture readings at a number of different points. When saving the individual
readings, the device will automatically calculate the readings’ average (see
"5.5
Saving several readings (measurement series) at the same time"
).
i
Information - Incorrect readings
Make sure to select the correct product type for the material you are measur-
ing. This prevents taking incorrect readings (see
"11. Faults"
).
i
Information - IR-temperature sensor
Pay attention to the IR temperature sensor on the top of your humimeter PMZ.
The temperature sensor is a precise optical measuring instrument. It should
not be under strong pressure when the electrodes are inserted or be damaged
by impacts. This will cause irreparable damage that falsifies the measurement
result because of incorrect temperature compensation.