User manual humimeter PMZ
Page 36
10.2 Calibration check
10.2.1 Calibration check of the lancet probe
Requirement:
The test block article no. 12308 is required.
The device and the test block have to reach a tempera-
ture between 23,0 °C and 23,5 °C.
The device and the test block must be free of dust, dirt
and grease.
1. Switch on the device and select the product type
"test block" using the arrow keys (see
6.1
).
2. Hold side 1 of the test block as shown on figure
61
on the electrodes.
»
The displayed water content has to be 22,0 %
(+/- 0,4 %) (the moisture value will be displayed in
black) (figure
62
).
»
If the displayed value lies outside this range (the
moisture value is displayed in grey) (figure
63
)
please, contact your dealer or Schaller Messtechnik
GmbH.
3. Hold side 2 of the test block as shown on
figure
64
on the electrodes.
»
The displayed water content has to be 41,0 %
(+/- 1,0 %) (the moisture value will be displayed in
black) (figure
65
).
»
If the displayed value lies outside this range (the
moisture value is displayed in grey) please, contact
your dealer or Schaller Messtechnik GmbH.
61
62
63
64
65