5.3 Initial start-up
Connect the network cable (6) to the power supply (5a). Connect the motor hand piece to the
'motor hand piece 1 'socket. Connect the controls satellite (3) to the control unit using the cable (9).
To switch on the device, activate the operating component (knee pad, foot lever, etc.) once.
If the worktop version is being used, the device is switched on by pushing the MENÜ
touchfield.
Plug the network adaptor into the power supply socket on the control unit.
.
Motorhandpiece 1
Socket for
control satellite
Power supply
Motorhandpiece 2
(only with
Premium)
Service socket
Foot pedal
Wax knife
Signal transmitter
for suction
(optional)
There is no main switch for the
control unit. If the hand piece is not used for more than 60 minutes,
it switches off automatically. Alternatively, the device can be switched off manually by simultaneously
depressing the touch-filed + for longer than 2 seconds. By simply touching the control device
(knee pad /foot pedal, MENÜ button on the worktop version), the device is switched on again.
6
In order to prevent unnecessary waiting times, when the wax knife is switched on, the control mechanism
is not automatically switched off.
Basically, firstly attach the handpiece, the controls satellite and the electric wax knife (if you have one)
before connecting the mains cable and switching the control mechanism on.
5.5 Connection of the signal transmitter for automatic swichting on of a workplace suction device
In order to operate the workplace system together with a workplace suction system with automatic
function, the signal transmitter for the suction signal must be connected. This is, as shown in the
illustration, inserted in the suction signal jack on the control unit. At the opposite end, in the appropriate
connector socket on the workplace suction device.
.
signal transmitter for suction
workplace suction with automatic
function
control unit
5.4 Switching on and off
When operation
together with a KaVo SMARTair suction or a Zubler suction AP 501, the optional cable Art.No.:9229
can be used as signal connection for controlling the suction.