CDRPanX User Guide
B1051407 Rev. A
25
6.3.
Using the Resend Option
The Resend option retrieves the last acquired image from the buffer inside the Sensor and
displays it in an empty viewbox that you select. This feature is appropriate when a
corrupted image is displayed on a system that, otherwise, appears to be operating
properly. In this instance, it may be possible to retrieve an uncorrupted version of the
image directly from the Sensor, without having to expose the patient to additional X-rays.
Figure 10. Resend Option Selected in Panoramic Settings
An example of how the Resend feature is used is provided below:
1.
After taking a panoramic X-ray, a corrupted image is displayed.
2.
IMPORTANT! Do not close CDR DICOM, EagleSoft, or Patterson Imaging as this
will clear the image in the Sensor.
3.
Locate the Resend option on the Panoramic device property page:
•
[CDR DICOM] System > Panoramic Settings.
•
[EagleSoft or Patterson Imaging] File > Preferences > X-ray > Schick Panoramic.
4.
Mark the “Retrieve last image from memory” checkbox and click OK.
5.
Click on an empty viewbox.
6.
Click Yes to the message, “Do you want to resend the last acquired image?”
7.
The image appears in the selected viewbox.
The Resend checkbox is cleared automatically after this function is used.