CDRPanX User Guide
B1051407 Rev. A
37
9. Maintenance
It is the responsibility of the user to maintain the equipment. Failure of the user to
properly maintain the equipment may relieve the manufacturer, or its agent, from
responsibility for any injury, damage, or non-compliance that may result.
9.1.
Visual Inspection
Like all electrical equipment, the CDRPanX system requires not only correct use, but
also visual inspection prior to operation, and routine checks at regular intervals. These
precautions will help ensure that the CDRPanX equipment operates accurately, safely,
and efficiently.
There are no user-serviceable components in the CDRPanX system. However, before
operating the system, users shall check it for any signs of physical damage or defect. If
detected, contact your local distributor of Schick Technologies products for further
instructions.
9.2.
Periodic Maintenance
Periodic maintenance is performed as needed, but at least once a month. It consists of
various checks performed by the operator or by a qualified service technician. Always
disconnect the main power supply before cleaning the system. Any defect or malfunction
should be corrected immediately by qualified personnel with adequate training. Tasks to
be performed regularly include the following:
•
To clean CDRPanX parts and surfaces, use isopropyl alcohol (wiped or sprayed),
or mild soap, taking care not to allow any liquid to seep into the
equipment. Plastic covers can be wiped with a soft cloth and light detergent.
•
To disinfect CDRPanX parts and surfaces that are in contact with the patient, use
a detergent (for example, one containing 2% solution of ammonia). The bite block
(but not the chin rest and the low and high chin templates) can be steam-sterilized
by autoclave at 250°F (121°C).
•
To disinfect CDRPanX parts and surfaces, use a solution of 70% ethanol as an
active ingredient. After the appropriate surfaces have been disinfected, they may
be wiped with isopropyl-soaked wipes / napkins.
•
Check all cables, any cables in deteriorated condition should be replaced.
WARNING: Any defective item affecting the safe use of CDRPanX
equipment must be repaired or replaced immediately. Refer to the
CDRPanX Service and Installation Guide for general service
information and specific procedures for repair and / or replacement.