Appendix A. Error Codes
Table 6. Control Panel Error Codes
Code Description
1 Microprocessor
error
2 EPROM
error
4 RAM
error
7
Rotation unit cannot move: stepper motor control error
11
Digital imaging software not ready to acquire image
12
Invalid limit switch status
13
Exposure exceeds maximum time permitted
14
X-ray high voltage error
15
Undervoltage condition at main power
17
Tube filament heating current exceeds range permitted
18
No current in X–ray tube
19
Exposure switch released during exposure time: incomplete image
20
Exposure switch released prior to X-ray emission: exposure was unsuccessful
21
Exposure switch released after X-ray emission, but rotation motor movement was not complete
23
Rotation unit steps do not match in half of the track
25
Rotation unit steps do not match in half of the track
28
Rotation unit steps do not match at the end of the track
31
Rotation unit does not move
33
Tube filament heating current active during off status
34
X–ray tube current active during off status
60
Control unit microcontroller or memory error
62
Software error while Rotation unit is in motion
63
Tube head temperature exceeds value permitted
64
15V power supply error
65
24V power supply error
66
5V power supply error
Please Note: The table above describes error codes that appear when the Error indicator on the Control Panel is