40
B1051407 Rev. A
CDRPanX User Guide
Appendix B. Technical Data
Table 7. Summary of CDRPanX Technical Data
Item Description
IEC Classification
Class I, type B
Power Supply
230V, 50/60 Hz, 8 A
115V, 50/60 Hz, 16 A
Mains Resistance
< 2 ohm at 230V
X-ray Generator
Multipulse at 20 kHz
Time Accuracy
+ 5%
kV Accuracy
+ 5%
X-ray Tube Current (mA)
+ 5%
Anode Voltage
60 - 86 kV, Constant Potential
Anode Current
4 - 10 mA, Direct Current
Focus Size
0.5 IEC 336
Inherent Filtration
0.098 in (2.5 mm) Al
Column Height
7.68 ft (234 cm)
Displacement
3.05 ft (93 cm), from 2.9 ft to 5.95 ft (88.5 to 181.5 cm)
Vertical Movement
Manual Adjustment
Patient Positioning
Manual Carriage Adjustment
Positioning Lasers
Lateral, mid-sagittal, frankfort
Centering Reference
Chin rest with bite stick
Focus Film Distance
1.67 ft (51 cm)
Exposure time
19 s
Projections
Adult (19 s)
Children (15 s)
Half Left (10 s)
Half Right (10 s)
Frontal Teeth (8 s)
TMJ opened and closed mouth (4 x 4 s)
Frontal Sinuses (19 s)
Cooling Pause
Automatically controlled, 6 minutes maximum
Weight
540 lbs (245 kg)
Table 8. Compliance Specifications
Item Description
CAN/CSA C22.2
No.601.1-M90
Medical Electrical Equipment Part 1: General Requirements for Safety
UL60601-1
Medical Electrical Equipment Part 1: General Requirements for Safety