CDRPanX User Guide
B1051407 Rev. A
41
Item Description
EC 93/42/EEC
Medical Device Directive
IEC60601-1
Medical Electrical Equipment Part 1: General Requirements for Safety
IEC60601-1-2
Medical Electrical Equipment Part 1: General Requirements for Safety
2.Collateral Standard: Electromagnetic Compatibility – Requirements and
Tests
EN 60601-1-3
Medical Electrical Equipment. Part 1: General requirements for safety.
3 - Collateral Standard: General requirements for radiation protection in
diagnostic X-ray equipment
EN 60601-2-7
Medical Electrical Equipment. Part 2: Particular requirements for the safety
of high voltage generators of diagnostic X-ray generators
EN 60601-2-28
Medical Electrical Equipment. Part 2: Particular requirements for the safety
of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
Table 9. Guidance and Manufacturer's Declaration - Electromagnetic Emissions
PLEASE NOTE: The CDRPanX system is intended for use in the electromagnetic environment
specified below. The customer or user of the CDRPanX system must ensure that it is used in such
an environment.
Emissions Test
Compliance
Electromagnetic Environment
RF emissions
CISPR 11
Group 1
The CDRPanX system uses energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Harmonic Emissions
IEC 61000-3-2
Not applicable
The CDRPanX system is suitable for use in domestic
establishments and in establishments directly connected
to the low-voltage power supply network that supplies
buildings used for domestic purposes.
Table 10. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
PLEASE NOTE: The CDRPanX system is intended for use in the electromagnetic environment
specified below. The customer or user of the CDRPanX system must ensure that it is used in such
an environment.
Immunity Test
IEC 60601-1-2 Test
Level
Compliance Level
Electromagnetic
Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%