USBCam
B1051039 Rev. B
33
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Guidance
< 70% U
T
(>30% dip in
U
T)
for 25
cycles
< 5% U
T
(>95% dip in
U
T)
for 5
sec
< 70% U
T
(>30% dip in
U
T)
for 25
cycles
< 5% U
T
(>95% dip in
U
T)
for 5
sec
NOTE: U
T
is the AC
mains voltage
prior to application of the test level.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3A/m 3A/m
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Portable and mobile RF
communication equipment should be
used no closer to any part of the
USBCam system, including its
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d= 1.2
√
P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d= 1.2
√
P for 80 MHz to 800Mhz
d= 2.3
√
P for 800 MHz to 2.5Ghz
Where P is the maximum output
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation in meters
(m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people