B1051061 Rev. A
USBCam2 User Guide
26
Table 6. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
PLEASE NOTE: The USBCam2 is intended for use in the electromagnetic environment specified
below. The customer or user of the USBCam2 must ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 610004-4
±
2 kV for power
supply lines
±
1 kV for
input/output lines
±
2 kV for power
supply lines
±
1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2kV common
mode
±
1 kV differential
mode
±
2kV common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5% U
T
(>95% dip in
U
T)
for 0.5
cycle
< 40% U
T
(>60% dip in
U
T)
for 5
cycles
< 70% U
T
(>30% dip in
U
T)
for 25
cycles
< 5% U
T
(>95% dip in
U
T)
for 5
sec
< 5% U
T
(>95% dip in
U
T)
for 0.5
cycle
< 40% U
T
(>60% dip in
U
T)
for 5
cycles
< 70% U
T
(>30% dip in
U
T)
for 25
cycles
< 5% U
T
(>95% dip in
U
T)
for 5
sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the USBCam2
requires continued operation during
mains interruptions, it is recommended
that the PC workstation to which the
USBCam2 is connected be powered from
an uninterruptible power supply or
battery.
NOTE: U
T
is the AC
mains voltage prior
to application of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m 3A/m
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
Portable and mobile RF communication
equipment should be used no closer to
any part of the USBCam2, including its
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d= 1.2
√
P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
3 V/m
d= 1.2
√
P for 80 MHz to 800Mhz
d= 2.3
√
P for 800 MHz to 2.5Ghz
Where P is the maximum output rating of
the transmitter in watts (W) according to
the transmitter manufacturer and d is the