B1051061 Rev. A
USBCam2 User Guide
iv
they are installed at a distance of at least 10 feet (3m) and supplied
by a dedicated electric line.
Apply Recommended Procedures for Cleaning the Equipment
Safe and proper operation of the equipment requires that a regular
schedule of preventive maintenance be followed. Refer to the
Protective Measures section of this manual for details.
Do Not Connect Items that are Not Part of the System
Only items specified for use with the equipment are to be connected
to the system. The equipment should not be used adjacent to other
equipment that is not part of the system. If, however, use with
adjacent equipment is necessary, normal operation should be
observed and verified in that configuration.
Installers to Ensure that USBCam2 Operates Optimally
Installers must ensure that the USBCam2 provides the user with the
optimal use of the equipment. This includes, but is not limited to,
ensuring the system operates as described in this document.
Installers must also ensure that the system presents no physical
obstacles or hazards during operation and when not in use. To verify
this requirement, installers shall confirm that the USBCam2 is
installed as described in this User Guide and shall perform the
appropriate procedures therein.
Ensure Proper System and PC Workstation Installation and Operation
The USBCam2 has been determined to be in accordance with
international safety standards and is deemed suitable for use within
the patient area, which extends from the patient for a distance of 5
feet (1.5m). To comply with these standards, do not operate non-
medical equipment (such as a PC workstation) inside the patient
area. Outside the patient area, the presence of approved non-medical
grade equipment and Listed / Approved / certified Information
technology Equipment (ITE) computer equipment is acceptable.
The PC workstation that connects to the USBCam2 via compatible
USB cable is an integral part of a Medical Electrical System. The
PC must be a CE-approved computer system conforming with the
Low Voltage [73/23/EC] and EMC Directive [89/336/ERC]. Also,
to help ensure optimal performance, make sure that all
software programs residing on the workstation are virus-free
and have been adequately tested so they will not impact imaging
applications after installation.
Please refer to documentation provided by the PC manufacturer for
important information about its safe operation and usage.