2
Components and Operation
2.2
Design
Page 16
A
rt.
no.
: 0-4
8-02
40 Rev. c
FRED PA-1
2.2.1
Overview of the configurable settings
Important!
Modifications that can be made via software program are only performed if re-
quested by the customer, or if required by legal requirements.
These modifications need to be registered in the device documentation as well
as communicated to all users.
SCHILLER's service centre can configure the following parameters:
Configurable parameters
• Selection of the default language at device start
• Energy level for 1st, 2nd and 3rd shock (separate settings for adults and children)
• Number of chest compressions for children (15 or 30)
• Self-test frequency (daily or weekly)
• Choice between "continuous chest compressions" or "alternating chest compres-
sions/breaths" during CPR cycles
• Date and time
• Update of the software/change of the device language
• Selection of the AED protocol (short or long instructions)
• Activation of notification if no RFID defibrillation pads are detected
• Activation of notch filter (50/60Hz)
• Activation of 16,7Hz Filter
• Activation of sound notification in case of elapsed maintenance interval
• Temperature warning
.