Initial operation
3.1
Inserting the battery
Page 22
A
rt.
no.
: 0-4
8-02
40 Rev. c
FRED PA-1
3.1.2
Switching device On and Off
Switching ON
Open the cover. The 3 LEDs for the resuscitation steps are briefly lit.
Switching OFF
Close the cover
Forced shutdown procedure
If the device cannot be switched off via the above procedure, remove the battery and
insert it again.
.
If a patient is detected while closing the cover, the device will remain ON and the
resuscitation process will go on.
If the cover is re-opened within 30 seconds after closing, the device will resume the
intervention.