Page 25
Initial operation
3
User Guide
Replacing the “pre-connected” pads
3.3
A
rt.
no.
: 0-4
8-02
40 Rev. c
FRED PA-1
3.3 Replacing the “pre-connected” pads
The
FRED PA-1
®
is delivered with "pre-connected" pads. To replaced the pads after
use or if the shelf life has expired, proceed according to the following instructions:
Only use the pads up to their expiration date.
Please note that the expiration date of the pads only applies if the vacuum pack
is intact.
Do not reuse the pads.
3.3.1
Connect the electrodes
1. Remove the battery
2. Remove the sticker with the LOT/Expiration date
from the electrode pouch
and stick it above the main status LED.
3. Open the cover.
4. Connect the electrode cable to the device.
5. Put the electrode pack in the cover and close the cover.
6. Make sure that neither the electrode cable nor the electrode packaging are
squeezed by the cover.
7. Insert the battery after closing the cover.
8. The device is ready for use when the status LED is blinking and the service LEDs
are off.
9. If requested, add a spare set of electrodes in the compartment on the device’s un-
derside.
Electrode connector