Page 39
Maintenance
6
User Guide
Maintenance Intervals
6.1
A
rt.
no.
: 0-4
8-02
40 Rev. c
FRED PA-1
6 Maintenance
6.1 Maintenance Intervals
• Because
FRED
®
PA-1
is
an emergency device, some verification has to
be done
as written in the following table
in
order to maintain the device operational, including
the accessories.The test results must be recorded and compared to the values ac-
companying the documents (see
• If used in optimal conditions (see Chapter
6.1.1 Exemption from the technical safe-
, the
FRED
®
PA-1
does not need any particular maintenance tests
since the device is able to test itself automatically on a regular basis, and it issues
a warning if any action either from the user or from a technician is required.
• Local regulations in your country may stipulate additional or different inspection in-
tervals and tests.
• The following table indicates the intervals and competence of the maintenance
work required.
Interval
Maintenance - replacement
Responsible
After each use
• Replace the electrodes.
• After battery insertion, check that the status LED is blinking and that
• Visual inspection of the device see
6.1.3 Visual inspection of the de-
User
Once a Week/Month
• Check that the green main status LED is blinking and all other LEDs
• Visual inspection of the device and accessories, see
spection of the device and accessories
.
User
Every 3 years
• Technical safety inspection is advised according to SCHILLER doc-
umentation (available for technical departments authorised by
SCHILLER), see
Note:
For an exemption from the 3-year technical service inspection, see
section
6.1.1 Exemption from the technical safety inspection
Service staff authorised
by SCHILLER
Every 6 years
• Replacement of internal backup battery. A technical safety inspec-
tion and a software update (if needed) are advised after opening the
device, see
Note:
The replacement of the internal backup battery is advised. Should this
internal backup battery not be replaced every 6 years, SCHILLER can-
not ensure the proper time stamping of the intervention.
Service staff authorised
by SCHILLER