Page 41
Maintenance
6
User Guide
Maintenance Intervals
6.1
A
rt.
no.
: 0-4
8
-02
40 Rev. c
FRED PA-1
6.1.3
Visual inspection of the device and accessories
Regularly and after each use, inspect visually the device and the cables in order to
detect possible mechanical damages.
If you observe damages or dysfunctions which can endanger the safety of the patient
or user, only use the device once it has been serviced.
Points to inspect:
• Check that the main status LED is blinking and all the other LEDs are off, see
• Device casing undamaged?
• No excessive clogging or damage?
• Legible nameplate at the rear of the device?
• Legible inscriptions on the front face of the device?
• Expiration date of the electrode elapsed? (see section
• Expiration date of the battery elapsed?
Electrodes past their expiration date must be replaced immediately (Electrode
and Service LEDs are on, only by using the electrodes reference 0-21-0040)
Batteries past their expiration date must be replaced immediately. (see expiring
date on the batteries)
Defective units or damaged cables must be replaced immediately.
Replace or repair immediately the device, if the main status LED is not blinking.
(see details in chapter