Page 51
Technical Data
7
User Guide
Classification and safety standards
7.2
A
rt.
no.
: 0-4
8-02
40 Rev. c
FRED PA-1
7.2 Classification and safety standards
Standards
FRED PA-1
®
complies with IEC standard 60601-2-4.
According to IEC standard 60601-2-4,
FRED PA-1
®
is a device for infrequent use.
EMC
.
Compliance
•
FRED PA-1
®
bears the
(Notified Body LNE/G-MED) mark indicating its
compliance with the provisions of the Directive 93/42/EEC (modified by the Direc-
tive 2007/47/EEC) regarding medical devices and fulfils the essential require-
ments of Annex I of this directive.
•
FRED PA-1
®
is a class IIb device.
Patient Protection
BF type, resistant to defibrillation shocks.
Explosions protection
FRED PA-1
®
is
not
designed to be used in the presence of flammable mixtures of
anaesthetic agents with air or oxygen.
The SCHILLER quality management system complies in full with the international
standards ISO 9001 and ISO 13485.