7.4 Electromagnetic interferences
The
FRED PA-1
®
is intended for use in the electromagnetic environment specified be-
low. The customer or the user of the
FRED PA-1
®
should assure that it is used in such
an environment.
7.4.1
Electromagnetic emissions
Emission measurement
Compliance with the regula-
tions
Electromagnetic environment - explanations
RF emissions
CISPR 11
Group 1
FRED PA-1
®
uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
FRED PA-1
®
is suitable for use in all establishments, including do-
mestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonics IEC 61000-3-2
Not applicable
Voltage fluctuations
IEC 61000-3-3
Not applicable
7.4.2
Electromagnetic immunity
Interference testing
IEC 60601 test level
Conformity level Electromagnetic environment - explanations
Electrostatic dis-
charge
IEC 61000-4-2
± 8 kV contact
± 15 kV air
IEC 60601-1
conformity
Floors should be made of wood, concrete or ceramic tiles. If floors
are covered with synthetic material, relative humidity should be at
least 30%.
Electrical fast tran-
sient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Not applicable
No mains power is used
Surge
IEC 61000-4-5
± 1 kV between conductors
± 2 kV conductor-earth
Not applicable
No mains power is used
Voltage dips, short in-
terruptions and volt-
age variations on
power supply input
lines
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
) for
0.5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cy-
cles
70 % U
T
(30 % dip in U
T
) for 25 cy-
cles
< 5 % U
T
(> 95 % dip in U
T
) for 5 s
Not applicable
No mains power is used
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
3 A/m
IEC 60601-1
conformity
Power frequency magnetic fields should be that of a typical com-
mercial and/or hospital environment.
Note: U
T
indicates the AC voltage of the mains before the test level.