7.7 Inspection report
The user guide must be read before the inspection.
Serial number
:
Checks - after each use
Check that the green indicator is blinking and all the other LEDs are off,
see
Visual inspection of the device and accessories
Device casing undamaged?
No excessive clogging or damage?
Legible nameplate at the rear of the device?
Legible inscriptions on the front face of the device?
Expiration date of the accessories elapsed?
Date:
Performed by:
Checks - once a Week/once a Month
Visual inspection of the device and accessories
(see previous table)
The main status indicator
1
is
lit green an no others LEDs are blinking
Date:
Performed by:
Checks
- every 3 years
Visual inspection of the device and accessories
(see previous table)
Functional test
Check for proper functioning (see
Measure the energy delivered at 50 ohms.
Date:
Performed by:
Replacement
-
every 6 years
Internal backup battery replacement.
Date:
Performed by:
In case of problems, please notify your Biomedical Department
, your local SCHILLER distributor
, or the authorized Custom-
er Service for your area
:
Name: .................................................................. Tel.: ...................................................................