31
as soon as a key is pressed or a weight is placed on the scale, BLOFF: backlight always off), press
to confirm, press
to escape.
10.2 Setting Auto Power Off
Hold the
key in for 3 seconds, display shows “setbl”. Press
key, display shows “setoff”.
Press
key to enter your selection, press
key to choose the auto power off time
(ON: always ON, 5: switch off auto after 5 mins. without operation, or 15: auto-switch off after
15 minutes without operation), press
key to confirm, press
key to go back.
10.3 Concerning a 4.0 BLE BT connection
Required password is "1234" to connect.
The default name to be found is "SCALE001".
11. Calibrating BWS computer
1.
Make sure that the animal cage weigher is level and flat with all its feet on the ground
2.
Make sure that the scale is empty and that nobody leans against it, as this affects the calibration
3.
Switch on the weighing computer press the on/off button
4.
Press the
button once during start-up (within 10 sec!)
5.
Display now shows PN
6.
Then press the following keys in sequence
,
and
7.
Display now shows PO CHT
8.
Then press the
button 3 times
9.
Display now shows P3 CAL
10.
Confirm with the
button
11.
Display now shows DECI
12.
Then press the
button 2 times until CAL is displayed
13.
Press
button again, NO LIN appears on the screen now
14.
Then press the
button, Display now shows UNLD
15.
Press
button again, weight becomes visible
16.
In this step you set the calibration weight you want to calibrate. With the
and
button you can set the decimals. With the
button you can change the decimal point.
17.
Then press the
button. Screen now shows load.
Summary of Contents for 4309606
Page 2: ...2 ...
Page 33: ...33 Pr 2 Send data to print and accumulation ...
Page 64: ...64 beschädigt ...