11
The available measuring range of the instrument can be reduced by the tare weight.
For example:
Measuring range
300 g
Minus tare (container etc.)
50 g
Remaining measuring range 250 g
Requirements:
- Instrument switched on as described in chapter 3.2.1.
Function settings selected as required, as described in chapters 3.3.1 and 3.3.2.
The ready symbol is displayed.
- Zero setting (auto zero) carried out as described in
chapter 3.3.3.
To measure with the tare function:
- Place the container on the platform.
- Press the
button, the display shows
.
NET appears under the ready symbol.
The tare has been subtracted from the measuring range.
The ready symbol must be shown on the display to ensure that the display
indicates correct readings.
To leave the tare function
:
- Remove the container from the platform and press the
button.
The display shows
.
- Press the
button.
The instrument is reset to the zero point (auto zero).
3.4.2 Measuring Procedure with Container
(tare subtractive weighing)
Ready symbol