4. Vacuum cushion by VakuForm.

interrupted, there is a risk that the flow of granules will stagnate and 
that the valve will become blocked. If this happens, you will have to 
briefly press air back from the transfer bag into the support system. The 
blockage should then clear quickly. If most of the air from the system 
has flowed into the transfer bag and the transfer procedure has not 
yet been completed, convey the air back into the cushion by pressing 
on the transfer bag. Then you can continue the transfer procedure as 
described above.

8. After completion of the transfer procedure, carefully pull the valve out 

of the rubber ring.

9. Now re-insert the closure, whereby it must be ensured that no more 

granules are located in the opening, as otherwise the tightness of the 
system may be adversely affected.

The granules are refilled analogue to the procedure described below in reverse.

Please leave the discharge of the granules to trained specialist personnel!

Summary of Contents for tina

Page 1: ...Instructions for use tina The cradle sub frame with cradle system tina...

Page 2: ...for use carefully prior to initial commissioning of the product and to observe them Please note that guidelines and representations in these Instructions for use may deviate from your product due to...

Page 3: ...5 Equipment for basic model 08 2 7 Emergency off switch 08 2 8 Product overview 09 3 Settings 10 3 1 Reclining surface height angle of inclination of head section 10 3 2 Leg angle 10 3 3 Seat angle 10...

Page 4: ...tem 20 6 3 Electrical data 21 6 4 Tyres 21 7 Guarantee 21 8 Identification 22 8 1 EU Declaration of Conformity 22 8 2 Serial number date of manufacture 24 8 3 Product version 24 8 4 Issue of the docum...

Page 5: ...eserve and protect the environment to prevent environmental pollution and to improve the recycling of raw materials please note the disposal instructions in points 1 3 1 and 1 3 2 1 3 1 Packaging The...

Page 6: ...nual control and the two electric motors are connected on the control unit with an emergency off switch and a battery Thebatteryisremovableandisrechargedviathechargerincludedin delivery which is mount...

Page 7: ...guide lines Correct usage of the product requires precise and careful training of the accompanying person The product may not be used without supervision Please observe Point 6 in these instructions f...

Page 8: ...n of the push handle Vacuum mattress Cover made from breathable fabric Support mattress Preparation of the sub frame for the assembly of a 12 E Fix drive Assemble the E Fix drive delivered Large suppo...

Page 9: ...nual operation brake lever for drum brake Battery pack control unit with emergency off switch Support mattress 8 PU steering wheel Vacuum mattress 2 Product description 2 8 Product overview The Fig be...

Page 10: ...the reclining surface Using the emergency off switch C you can suppress the manual control function against inadvertent activation 3 2 Leg angle The leg section on the three section crad le system is...

Page 11: ...Overcharging is continued charging of the battery pack after it has already been fully charged Permanent overcharging shortens the lifetime of the battery pack In order to achieve 100 charging a char...

Page 12: ...ly on one side it is possible to achieve a steering movement when pushing 3 7 Push bar The push bar can be moved down after release of the two star handle screws C Then re tighten the star handle scre...

Page 13: ...wedge shape As soon as the required sinking depth is achieved you can start with correc tion work The aim is to achieve as symmetrical a position as possible within the scope of the illness symptoms...

Page 14: ...te trans fer spout A is located at the uppermost point of the system 2 Add enough air via the valve B into the system until the granules slide downwards and the spout is lifted clear 3 The closure is...

Page 15: ...as not yet been completed convey the air back into the cushion by pressing on the transfer bag Then you can continue the transfer procedure as described above 8 After completion of the transfer proced...

Page 16: ...age in a closed vehicle or long direct solar radia tion may lead to softening of the support system for physical reasons In extreme cases this may lead to a loss of dimensional stability Such a loss o...

Page 17: ...evenly applied with a soft cloth Alternatively wipes pre soaked with disinfectant can be used to wipe the products over the entire surface In both cases care must be taken to ensure complete wetting...

Page 18: ...ectrical height adjustment back adjustment and adjustment of angle of inclination including the wiring Functionality of the vacuum mattress and the related electrical pump Functionality of brakes Func...

Page 19: ...iod if it is in a safe condition The expected duration of use does not refer to wear parts such as for example wheels gas pressure springs The maintenance and evaluation of the condition and if applic...

Page 20: ...steplessly adjustable 0 18 D Adjustable foot angle 0 50 E Tilt angle reclining surface 0 20 F Length of cradle support 95 cm Base frame width 64 cm 64 cm Weight 44 kg 44 kg max load 100 kg 100 kg A D...

Page 21: ...min continuous operation 18 min break Protection class IP 66 Electric motor Operating voltage 24 V DC back adjustment Max Force 4000 N Duty cycle max 2 min continuous operation 18 min break Protection...

Page 22: ...listed directives and standards Verordnung EU 2017 745 ber Medizinprodukte vom 05 April 2017 Regulation EU 2017 745 on medical devices of 5 April 2017 DIN EN 12182 2012 Technische Hilfen f r behinder...