page 14
Fault
causes
(see also Figure 1 on page 7)
Possible causes:
Remedy:
No contact water
Fill transducer head (6) up to the
mark with distilled water
Transducer head incorrectly
inserted
Fit transducer head (6) up to the
limit stop. Watch out for the anti-
rotation device at the back
In the event of poor aerosol flow due to
Air bubbles forming on
the quartz or in the nebulizing
chamber (5)
Switch off the nebulizer, remove
bubbles
Moist bacteria filter (11)
Exchange filter. Discontinue use of
wet filter!
Accumulation of water in the
sagging hose (12)
Reposition the hose with a
continuous slope
Water level in chamber (5) too
high
Reduce the water level
If malfunctions cannot be remedied, contact the authorized service
technician or your dealer.
The manufacturer will also provide on request circuit diagrams, spare
parts lists, descriptions and adjustment instructions for the U-3003-E
Ultrasonic Nebulizer, which the user’s qualified personnel can use to
repair the nebulizer.