SPORÉCUP XTR2   |   Instruction Manual 

Intended use 

The XTR2 has been designed as a device for the generation of transcutaneous electrical 
stimulation of the nerves and muscles on humans and may not be used for any other purposes. 

Please read the operating instructions carefully before using the XTR2. 

 

Safety instructions 

 

Important safety 

The XTR2 is only to be used for stimulation purposes on one patient at any one time.  
Only use the XTR2 with original accessories. The size of the electrodes should to be not less 
than 2 cm².The electrodes should be positioned in such ways that the current flow will not cross 
through the heart. 
Keep the XTR2 away from water or other liquids. Do not drop the XTR2, do not use it 
inappropriately or expose it to extreme temperatures or high levels of humidity (not less than 
10°C or more than 40°C or a relative humidity of more than 90 %).  
Do not use the XTR2 if it is not working properly or if it has been damaged in any way.  
Be careful when using the XTR2 near or on children. Keep the unit away from children. 
Always store the XTR2 in its case to protect it from damage and dust. 
To avoid reciprocal interference, the XTR2 should not be operated in the vicinity of other 
electronic devices. If this is not possible, the various unit functions should be closely monitored 
during operation so that proper use in accordance with the regulations is ensured. The 
simultaneous connection to the patient of a high frequency surgical unit can lead to burning under 
the electrodes or around the probe. Operating close (for example 1m) to short wave and 
microwave units or mobile telephone systems can cause fluctuations in the output values of the 
electric current stimulation unit. 
Do not use the machine during the simultaneous operation of other machines, when driving or 
when asleep. 
 

Contra-indication 

Please check with your physician before using the XTR2 if one of these applies to you: 

  Patients with electronic implants such as heart pacemakers, defibrillators or pumps 

  Patients with cardiac rhythm disorders 

 During 

pregnancy 

  Patients subject to seizure disorder / epilepsy 

  Patients with skin disorders in the vicinity of the electrodes 

  Patients with malignant diseases in the region of application 

 

Side effects 

  Pain caused by the stimulation: The stimulation can be perceived as uncomfortable/painful if 

the electric current is too intensive or if the electrodes have been positioned unfavorably. 
Unpleasant side effects can be avoided by adjusting the intensity, using another program with 
other parameters and the possible re-positioning of the electrodes. 

  Skin intolerance which can result from the electrodes, the electrode gel or the electric current 

impulses. The physician should be consulted in cases of reddening, burning, itching or 
blistering under the electrodes or in the vicinity of the electrodes. Slight skin reddening of 
short duration in the area of the electrodes following stimulation is quite normal because 
blood circulation has been improved by the effects of the stimulation. 

  Muscular pain: If the stimulation has been too intense or too long muscular aches in terms of 

sore muscle can occur. Shorten the treatment time and the intensity of stimulation at the 
beginning of muscle stimulation to avoid muscle ache. 
 

Summary of Contents for 104064

Page 1: ...SPORÉCUP XTR2 MUSCLE STIMULATOR FOR SPORTS AND FITNESS Instruction Manual Art no 101478 ...

Page 2: ...r 7 Switching OFF the stimulator 7 Charging the battery 7 Description of the programs 8 7 Setting of intensity levels 10 8 General information 10 9 Delivery Content 11 10 Accessories 11 Stimex self adhesive electrodes 12 Stimex FLOWER POWER self adhesive electrodes 12 Silicone electrodes to be used with contact gel 12 Contact gel for silicone electrodes item no 108000 Tube 60 g Fehler Textmarke ni...

Page 3: ...erating close for example 1m to short wave and microwave units or mobile telephone systems can cause fluctuations in the output values of the electric current stimulation unit Do not use the machine during the simultaneous operation of other machines when driving or when asleep Contra indication Please check with your physician before using the XTR2 if one of these applies to you Patients with ele...

Page 4: ... instructions manual 0197 Conformation with Council Directive 93 42 EEC of 14 June 1993 concerning medical devices BF type application part Protection against electric shocks SN The serial number of the product follows this symbol REF The article or order number of the product follows this symbol This equipment is marked with the recycling symbol It means that at the end of the working life of the...

Page 5: ...ut short circuit control element AKS 30 integrated pre set programs Output current 100mA with an actual resistor of 1kΩ Frequency range 0 5 120 Hz Impulse width 50 400 μs Nominal current 15 mA Power supply integrated rechargeable accumulators Ni mH 4 8 V Dimensions 11 4 x 5 9 x 2 7 cm Weight approx 300 g 5 SPORÉCUP XTR2 Instruction Manual ...

Page 6: ...chapter interlocking the key pad Setting the intensity Start the stimulation by increasing the intensity with the key of the channel s connected to the electrodes key to increase the intensity key to decrease the intensity The intensity of each channel is displayed on the screen The intensity can be set between 0 and 100 mA By keeping the intensity keys pressed the intensity will increase decrease...

Page 7: ...omatically Charging the battery The battery level is shown on the display with a battery symbol with 4 lines If the battery level is too low the stimulator switches off automatically and cannot be turned on again The battery needs to be re charged Switch the commutator to the OFF position no 8 Connect the charger with the XTR2 socket no 7 Connect the charger to the mains circuit The diode on the c...

Page 8: ...sms P2 Recovery after competition This program uses even a lower frequency than P1 in order to distress the tired muscles The blood flow is accelerated while the activity automatically slows down and the muscle relaxation is intensified The muscles recover better and faster for optimized future performance P3 Recovery after force training This program is recommended after an intensive force traini...

Page 9: ...atica P14 Tendinitis P15 Joint pain P16 Chronic pain endorphinic SPORT P17 Resistance P18 Endurance P19 Strength P20 Explosive strength P21 Hypertrophy body building REHABILITATION P22 Amyotrophy prevention CAPILLARIZATION P23 Capillarization 1st phase P24 Capillarization 2nd phase FITNESS P25 Muscle starter P26 Toning muscle building ...

Page 10: ...mall muscles Muscle fatigue A tired muscle will only tolerate a lower intensity Age of electrodes Self adhesive electrodes do not last forever You can use them up for up to 60 stimulation sessions This depends on the quality of the electrodes but also on the skin of the user Old electrodes have a higher resistance and their conductivity levels decline It is important to replace electrodes regularl...

Page 11: ...n the XTR2 every 24 months This includes 1 Confirming that user instructions have been included in the accompanying documentation 2 Checking the equipment for completeness 3 Visual check for mechanical damage for damage to all cables and sockets 4 Functional Safety Checking the output signals with a load resistance of 1 kΩ real current and voltage Checking the frequency Checking the pulse width Th...

Page 12: ... 283600 Stimex 50x90 mm 2 281007 Stimex 50x90 mm 4 283000 Stimex 50x130 mm 2 283100 Stimex 80x130 mm 2 281027 Stimex sensitive 50x50 mm 4 Stimex FLOWER POWER self adhesive electrodes Art No Description Quantity 281042 Flower Power 50 mm round 4 281035 Flower Power 50x50 mm 4 281040 Flower Power 50x90 mm 4 Silicone electrodes to be used with contact gel Art No Size Quantity 107090 20 mm round 2 107...

Page 13: ...age with pelotte and integrated electrodes to be used with the XTR2 MedicoBack P type offers patients lumbar vertebrae stabilisation and support as well as easy nerve stimulation thanks to the integrated electrodes The pelotte with its massage nubs intensifies the pain relieving effect MedicoBack P type may be used for light or acute pain as well as for chronic pain in the lower back area Art No A...

Page 14: ...12 Electrode placement Electrode placement for pain treatment Back pain Shoulder pain Epicondylitis ulnaris Epicondylitis radialis 14 SPORÉCUP XTR2 Instruction Manual ...

Page 15: ...Arthrosis hip Kaada one canne only Sciatic pain Gonarthrosis Knee pain Ankle joint pain Trigeminus neuralgia 15 SPORÉCUP XTR2 Instruction Manual ...

Page 16: ...16 SPORÉCUP XTR2 Instruction Manual Migraine Tension headache Polyarthritis Polyneuropahtia Carpal tunnel syndrome Pain of achilles tendon heel pain ...

Page 17: ...requency parameters for an improved pain control or with low frequency parameters for optimized muscle stimulation This treatment method also has an enhanced effect on lymphatic drainage stimulation The placement of electrodes for dynamic stimulation depends on the direction of stimulation rising falling stimulation or sideways stimulation Rising or falling stimulation Sideways stimulation for bac...

Page 18: ...Electrode placement for muscle stimulation Hand flebo Hand extensor Biceps Back Quadriceps Leg flebo hamstrings Foot flexor M tibialis Calf 18 SPORÉCUP XTR2 Instruction Manual ...

Page 19: ...y building P6 Next day regeneration P7 Venous return dynamic stimulation REHABILITATION P8 Heavy legs P22 Amyotrophy prevention RELAXATION CAPILLARIZATION P10 Slow dynamic massage P23 Capillarization 1st phase P24 Capillarization 2nd phase PAIN TREATMENT FITNESS P11 Cervical pain P25 Muscle starter P12 Low back pain P26 Toning muscle building P13 Lumbosciatica P14 Tendinitis P15 Joint pain P16 Chr...

Page 20: ...r GmbH Hoersnheimer Eck 19 35578 Wetzlar Germany www pierenkemper eu Distributer schwa medico GmbH Wetzlarerstr 41 43 35630 Ehringshausen Germany www schwa medico com 20 SPORÉCUP XTR2 Instruction Manual export schwa medico de ...

Reviews: