Program No.
Program name
Indication
Description
Frequency/Hz
Pulse width/µs
Time/min
1
Gate control 1
Nociceptive pain, acute pain, neuropatic pain
Both channels
100
200
30
2
Gate control 2
Alternative to program 1
Both channels
80
150
30
3
Low frequency
Chronic nociceptive pain, improvement of
blood circulation, Kaada stimulation
Both channels
2
250
30
4
Gate c
low frequency
Combination of high and low frequency stimulation
Channel 1: 100 Hz
Channel 2: 2 Hz
Channel 1: 100
Channel 2: 2
200
30
5
Adaptation
Muscular pain
First 10 min 100 Hz, then 20 min with 2 Hz
100/2
150/200
10/20
6
HAN
Suitable for almost all kinds of pain program
3 s 100 Hz, then 3 s 2 Hz etc.
100/2
150/200
30
7
Burst
Alternative to program 3 (more comfortable)
100 Hz for 0.25 s
100
150
30
8
Modulation
Alternative to the other programs in case
of therapy resistance
Automatic frequency run from
2 Hz to 80 Hz within 7.5 s
2 - 80
200 - 100
30
9
Muscle stimulation
Atrophy prophylaxis, muscle strengthening
Increase 2 s, working time 5 s, decrease 1 s,
pause 12 s, automatic intensity adaption
50
250
30
10
Gate control
dynamic 1
Muscle relaxation
Intensity within 1 s to maximum,
then within 1 s reduced to 50%.
Both channels in alternating operation
80
150
30
11
Gate control
dynamic 2
Muscle relaxation, venous return
Intensity within 0.25 s to maximum,
then within 0.25 s reduced to 50%.
Both channels in alternating operation
80
150
30
12
Deep TENS
Muscular pain, deep seatted pain
Impuls packages of 4 impulses.
200 µs between each impuls
100
75
30
User programs
13
TENS stimulation
Pain treatment (see programs 1-3)
Both channels
1-120
75-300
10-90
14
Muscle stimulation
Muscular pain, deep seatted pain
Increase 2 s, working time 5 s, decrease 1 s,
pause 12 s (automatic intensity adaptation)
1-120
75-300
10-90
15
Gate control dynamic
Muscle relaxation
Intensity within 1 s to maximum, then within 1 s
reduced to 50%. Both channels in alternating operation
1-120
75-300
10-90
16
Monophasic
Iontophoresis, nerv block, Jenkner block
(stellate ganglion block)
Both channels with monophasic current
1-120
75-300
20