Bravo G4 17 / Bravo G4 22 / Bravo G4 28
EN
25
5.2. MAIN MENU
At the end of starting procedure the main menu is displayed on the side.
The steriliser waits for the program selection (see "Program selection"
Chapter).
5.3. FILLING DEMINERALISED / DISTILLED WATER
Only use demineralised / distilled water having the following characteristics to supply the steriliser.
DESCRIPTION
VALUES IN THE WATER SUPPLY
VALUES INSIDE RESIDUAL
DRY CONDENSATE
< 10 mg/l
< 1 mg/l
SILICONE OXIDE SiO
2
< 1 mg/l
< 0.1 mg/l
IRON
< 0.2 mg/l l
< 0.1 mg/
CADMIUM
< 0.005 mg/l
< 0.005 mg/l
LEAD
< 0.05 mg/l
< 0.05 mg/l
HEAVY METAL RESIDUES (iron, cadmium
and lead excluded)
< 0.1 mg/l
< 0.1 mg/l
CHLORIDES
< 2 mg/l
< 0.1 mg/l
PHOSPHATES
< 0.5 mg/l
< 0.1 mg/l
CONDUCTIVITY AT 20°C
< 15 µS/cm
< 3 µS/cm
pH VALUE
5 - 7
5 - 7
ASPECT
colourless, transparent, without sediment
colourless, transparent, without sediment
HARDNESS
< 0.02 mmol/l
< 0.02 mmol/l
5.3.1.
MANUAL FILLING
The first time the steriliser is used, and later when the lack of water alert
is signalled, you will have to fill, or top-up, the internal clean water tank.
Remove the cap from the top cover.
Pour in water taking care not to exceed the maximum level indicated
inside the tank (MAX.). Close the door.
Pay attention not to spill water on the machine; if water is spilled, promptly
dry.
The tank must be filled before the cycle starts or after its completion.
Do not open the tank doors during the cycle execution in order to prevent hot water leaks.
5.3.2.
AUTOMATIC FILLING
Refer to the “ACCESSORIES” appendix.