Bravo G4 17 / Bravo G4 22 / Bravo G4 28
74
EN
PROBLEM
POSSIBLE CAUSE
SUGGESTED SOLUTION
Contact between instruments made of different
metals.
Separate instruments made of different metals.
(See
Chapter “Preparing the material”).
Presence of limescale residues on the wall of the
chamber and/or accessories.
Clean the chamber and the accessories as prescribed.
(See
Appendix “Maintenance”).
Blackening of the instruments or
damage to the material.
Wrong selection of the sterilisation program.
Choose the sterilisation program suitable for the type
of material to be treated.
(See the
Summary Table in "Programs" Appendix).