Bravo G4 17 / Bravo G4 22 / Bravo G4 28
78
EN
17.2. ALARMS (CATEGORY A)
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET MODE
A032
Problem with the level sensor of the filling tank
FILL.WATER LEVEL
SENSOR PROBLEM
CONTACT TECH.SERVICE
1
A033
Problem with the level sensor of the discharge tank
DISCH.WATER LEVEL
SENSOR PROBLEM
CONTACT TECH.SERVICE
1
A040
The tank has not been filled (only with automatic filling system)
FAILED WATER INLET
CHECK
AUTOMATIC LOAD
1
A042
The MAX. water level in the filling tank has been reached abnormally
(automatic filling)
WATER FILLING
MAXIMUM LEVEL
CHECK TANK
1
A043
Discharge tank full beyond maximum limit
DISCHARGE TANK
MAXIMUM LEVEL
CHECK TANK
1
A101
Temperature sensor PT1 broken
(sterilisation chamber)
CHAMBER PROBE
PT1 OPEN CIRCUIT
CONTACT TECH.SERVICE
1
A102
Temperature sensor PT2 broken (steam generator)
GENERATOR PROBE
PT2 OPEN CIRCUIT
CONTACT TECH.SERVICE
1
A103
Temperature sensor PT3 broken (heating element)
HEATING BAND PROBE
PT3 OPEN CIRCUIT
CONTACT TECH.SERVICE
1
A105
Temperature sensor PT5 broken (conductivity measurement
compensation)
CONDUCTIVITY SENSOR
PT5 OPEN CIRCUIT
CONTACT TECH.SERVICE
1
A111
Temperature sensor PT1 short-circuited (sterilisation chamber)
CHAMBER PROBE
PT1 SHORT-CIRCUIT
CONTACT TECH.SERVICE
1
A112
Temperature sensor PT2 short-circuited (steam generator)
GENERATOR PROBE
PT2 SHORT-CIRCUIT
CONTACT TECH.SERVICE
1
A113
Temperature sensor PT3 short-circuited (heating element)
HEATING BAND PROBE
PT3 SHORT-CIRCUIT
CONTACT TECH.SERVICE
1
A115
Temperature sensor PT5 short-circuited (conductivity measurement
compensation)
CONDUCTIVITY SENSOR
PT5 SHORT-CIRCUIT
CONTACT TECH.SERVICE
1
A116
ADC error
PROCESS BOARD ERROR
CONTACT TECH.SERVICE
1
A120
Reference heating element acquisition chain fault
PROCESS BOARD ERROR
CONTACT TECH.SERVICE
1
A121
Reference heating element acquisition chain fault
PROCESS BOARD ERROR
CONTACT TECH.SERVICE
1
A122
Reference heating element acquisition chain fault
PROCESS BOARD ERROR
CONTACT TECH.SERVICE
1
A123
Solenoid valves control driver fault
EV DRIVER ERROR
CONTACT TECH.SERVICE
2
A124
Door motor control driver fault
DOOR MOTOR
DRIVER ERROR
CONTACT TECH.SERVICE
2
A125
Faulty current draw detected
FAULTY
CURRENT DRAW
CONTACT TECH.SERVICE
2
A126
Connection error with WiFi module
WI-FI MODULE ERROR
CONTACT TECH.SERVICE
1
A128
Communication error between graphical interface and ethernet module
ETHERNET MODULE
ERROR
CONTACT TECH.SERVICE
1
A131
Solenoid valve 1 failed
SOLENOID VALVE 1 ERROR
CONTACT TECH.SERVICE
2
A132
Solenoid valve 2 failed
SOLENOID VALVE 2 ERROR
CONTACT TECH.SERVICE
2
A133
Solenoid valve 3 failed
SOLENOID VALVE 3 ERROR
CONTACT TECH.SERVICE
2
A134
Solenoid valve 4 failed
SOLENOID VALVE 4 ERROR
CONTACT TECH.SERVICE