17.5.4. SYSTEM ERRORS (CATEGORY S)
CODE
POSSIBLE CAUSE
SUGGESTED SOLUTION
S001
Error of Flash memory 1 on process board
Flash memory 1 on process board failed
Contact Technical Service (see Appendix).
S002
Error of Flash memory 2 on process board
Flash memory 2 on process board failed
Contact Technical Service (see Appendix).
S003
Error of flash memory 1 on graphical interface board
Flash memory 1 on graphical interface board failed
Contact Technical Service (see Appendix).
S004
Error of flash memory 2 on graphical interface board
Flash memory 1 on graphical interface board failed
Contact Technical Service (see Appendix).
S005
USB key not correctly formatted
Damaged USB key
Check USB key correct formatting (FAT32).
As an alternative, use another correctly formatted USB key.
If the problem persists, contact Technical Service (see Appendix).
S006
USB key not correctly formatted
Damaged USB key
Check USB key correct formatting (FAT32).
As an alternative, use another correctly formatted USB key.
If the problem persists, contact Technical Service (see Appendix).
S007
USB key full
Download data from USB key or use another USB key.
If the problem persists, contact Technical Service (see Appendix).
S009
Printer off.
Data cable not correctly connected to serial ports RS-
232.
Make sure that printer is on.
Check printer cable correct connection.
If the problem persists, contact Technical Service (see Appendix).
S010
No paper inside printer.
Paper setting configuration not correctly done.
Make sure that paper is correctly loaded.
Check printer cable correct connection.
Make sure that paper settings are correct.
If the problem persists, contact Technical Service (see Appendix).
S011
Printer lid open
Make sure that printer lid is correctly closed.
Check printer cable correct connection.
If the problem persists, contact Technical Service (see Appendix).
S012
Printer not ready for use
Make sure that paper is correctly loaded.
Check printer cable correct connection.
Make sure that paper settings are correct.
If the problem persists, contact Technical Service (see Appendix).
S020
Cycle back-up not done after 250 cycles
Perform cycle back-up.
See paragraph Sterilisation cycle back-up.
If the problem persists, contact Technical Service (see Appendix).
S021
Cycle storage limit exceeded after 500 cycles
Perform cycle back-up.
See paragraph Sterilisation cycle back-up.
If the problem persists, contact Technical Service (see Appendix).
S030
Malfunction of the control software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service (see the Appendix).
S031
Malfunction of control board or software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service (see the Appendix).
S032
Malfunction of the control software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service (see the Appendix).
S034
Malfunction of the control software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service (see the Appendix).
S035
Control software malfunction in solenoid valve
management
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service (see the Appendix).
S040
Malfunction of the control software
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service (see the Appendix).
S041
Malfunction of control board or
control software
Contact Technical Service
(see Appendix).
S042
Malfunction of control board or software
Contact Technical Service (see Appendix).
S099
Malfunction of control board or
control software
Try restarting the program a second time.
Try replacing the USB key.
If the problem persists, contact Technical Service (see the Appendix).
S100
Malfunction of control board or
control software
Contact Technical Service (see Appendix).