Bravo G4 17 / Bravo G4 22 / Bravo G4 28
EN
11
3.2. TECHNICAL SPECIFICATIONS
3.2.1.
SUMMARY TABLE
Device
STEAM STERILISER
Bravo G4 17
Bravo G4 22
Bravo G4 28
Class (according to Directive
93/42/EEC and subsequent
amendments)
IIb
Manufacturer
CEFLA s.c.
Sede legale - Headquarter
Via Selice Provinciale 23/A
– 40026 Imola (BO) IT
Input voltage
220 V - 240 V~ 50 Hz
220 V - 240 V~ 60 Hz
Network fuses
(6.3 x 32 mm)
2x T15A 250V
Electronic board fuses
(5 x 20 mm)
F1: T3.15A 250V (transformer primary 220/240 V~ 50 Hz 220/240 V~ 60 Hz)
Nominal power
2300 W
Insulation class
Class I
Installation category
(according to EN 61010)
Cat. II
Operational environment
Indoor use
HUMID LOCATION
(EN 61010 extended environmental conditions)
A-weighted sound power level
(ISO 3746)
< 67 db (A)
Degree of protection (IP Code) (EN
60529:1991+A1:2000+A2:2013)
IP21
Environmental operating conditions
Temperature: +15°C ÷ +35°C
Relative humidity: between 20% and 80% max. non-condensing
External dimensions (HxWxD)
(rear connections excluded)
450 x 480 x 600 mm
Net weight:
unladen
approx. 48 kg
approx. 49 kg
approx. 50 kg
unladen, with tray holder support
and trays
approx. 51 kg
approx. 52 kg
approx. 53 kg
unladen, with tray holder support,
trays and water at MAX. level
approx. 53 kg
approx. 54 kg
approx. 55 kg
Sterilisation chamber dimensions
(D x D)
250 x 350 mm
250 x 450 mm
280 x 450 mm
Sterilisation chamber total volume
approx. 17 l
(0.017 cu. m)
approx. 22 l
(0.022 cu. m)
approx. 28 l
(0.028 cu. m)
Sterilisation chamber usable
volume (with tray holder support
inserted)
approx. 10 l
(0.010 cu. m)
approx. 13 l
(0.013 cu. m)
approx. 19 l
(0.019 cu. m)
Sterilisation chamber usable
dimensions
17 l (1.38x1.55x2.97) dm /
6.4 cu. dm
22 l (1.38x1.55x3.97) dm /
8.5 cu. dm
28 l (1.72x1.66x3.96) dm /
11.3 cu. dm
Water tank capacity (filling)
approx. 5.5 l (water at MAX. level)
approx. 1 l (water at MIN. level)
Sterilisation programs
5 standard pr 1 program defined by the user
Test programs
Helix/B&D Test
Vacuum Test
Vacuum Test+Helix/B&D Test
Pre-heating time
(from cold)
approx. 10 min.
USB connection
Key capacity lower than or equal to 2GB: FAT formatting with 16K/sector
Keys capacity higher than 2GB:
FAT32 formatting with 16K/sector
Printer connection
Serial RS232 (printer cable max. length 2.5 m)
Printer insulation class:
Class I or Class II