Bravo G4 17 / Bravo G4 22 / Bravo G4 28
EN
41
8.3. EXECUTION OF THE CYCLE
Taking as an example the most complete and significant sterilisation
cycle, i.e. the
B 134°C UNIVERSAL
program, characterised by
fractionated pre-vacuum, the cycle sequence is as follows:
WARMING UP
FIRST VACUUM PHASE
FIRST PRESSURE RISE
SECOND VACUUM PHASE
SECOND PRESSURE RISE
THIRD VACUUM PHASE
THIRD PRESSURE RISE
STERILISATION
STEAM DISCHARGE
DRYING
VENTILATION
CYCLE COMPLETION
8.4. CYCLE OUTCOME
At the end of the cycle it is important to check the sterilisation process
outcome.
If the message "
COMPLETED
" is displayed, it means that the cycle has
been completed correctly without any interruptions for alarms and that
complete asepsis
of the material is guaranteed.
8.5. DOOR OPENING AT THE END OF THE CYCLE
To open the steriliser door, press the button shown in the figure: