14.2. SCHEDULED MAINTENANCE MESSAGES
The steriliser periodically displays warning messages relevant to "routine" maintenance operations that must be carried out in order to ensure the proper
operation of the device.
Press DONE to confirm that the required maintenance operation has been
completed.
Press REMIND button to postpone the operation.
In this case, the warning message will reappear the next time the steriliser
is used.
Always keep in mind the following
general warnings:
•
Do not
wash the steriliser with direct jets of water, neither under pressure nor sprinkled. Seepage into electrical and electronic components could
damage the functioning of the device or its internal parts, even irreparably;
•
Do not
use abrasive cloths, metal brushes (or other aggressive materials) or products for metal cleaning, both solid and liquid, to clean the device or
the sterilisation chamber;
•
Do not
use unsuitable chemical products or disinfectants to clean the sterilisation chamber. In fact, these products can cause damage, even irreparably;
•
Do not
allow limescale or residues of other substances to accumulate in the sterilisation chamber, on the door and on the gasket, but remove them
periodically. In fact, such residues may damage these parts, besides compromising the operation of the hydraulic circuit components.
The formation of white spots on the base of the internal wall of the chamber means that you are using poor quality demineralised water.
Before performing regular maintenance, make sure that the
power cord plug is removed from the mains socket.
If this is not possible, move the external switch of the
device’s power supply line to Off.
If the external switch is distant or not visible to the
maintainer, place a “work in progress” sign on the switch,
after turning it off.