Bravo G4 17 / Bravo G4 22 / Bravo G4 28
EN
71
14.3.6. ANTI-DUST FILTER CLEANING
Remove the filter (
A
) from the lower part of the autoclave, thoroughly rinse
it with water and dry it before refitting it.
The filter can be cleaned using a jet of compressed air, making sure not
to disperse any dust into the environment.
14.3.7. REPLACE THE BACTERIOLOGICAL FILTER
When filter maintenance is due or every time you notice visible clogging
of the filter (indicated by the filter markedly turning grey), unscrew the
bacteriological filter (
B
) from its support and replace it with a new one,
screwing it fully down on the union.
A spare bacteriological filter is provided with the device.
If you need spare parts of this component, refer to technical
assistance appendix.
14.3.8. BOILER GASKET REPLACEMENT
The boiler gasket needs to be replaced by an authorised technician.
Contact Technical Service (see
APPENDIX
– TECHNICAL SERVICE
).