17.5. TROUBLESHOOTING
According to the
type of alarm
occurred, please find below the indications to detect the possible causes and restore the proper operation:
17.5.1. ERRORS (CATEGORY E)
The alarm codes in the list can refer to functions that are not present on the models concerned in this Manual
CODE
POSSIBLE CAUSE
SUGGESTED SOLUTION
E000
Sudden power failure (blackout).
Wait for the power to be restored and do a RESET following the instructions.
The main switch has accidentally been turned off
and/or the power plug pulled from the socket.
Reconnect the plug and/or turn the device on again and RESET according to the
instructions.
Network fuses blown.
Contact Technical Service.
E001
Abnormal voltage peak on the mains.
Reset according to the instructions. If the problem occurs again, have the mains
electric system checked by a technician.
E002
The filling tank contains water of inadequate quality.
RESET according to the instructions. Empty the filling tank and refill it with
demineralised /
distilled water of adequate quality (<15μs/cm).
E003
The filling tank contains water of very poor quality.
RESET according to the instructions. IMMEDIATELY empty the filling tank and
refill it with demineralised /
distilled water of adequate quality (<15μs/cm).
In these conditions, the steriliser allows starting a maximum of 5
cycles, after which it locks until the tank is filled with demineralised /
distilled water of adequate quality (<15 μs/cm). This precaution is
necessary to prevent damage to the device.
E004
Failure to main board.
RESET according to the instructions.
Contact Technical Service (see Appendix).
Disturbance on the electrical mains.
RESET according to the instructions.
If the problem occurs again, have the electrical mains checked by a technician.
If the electrical mains is equipped with a Continuity system, have the system
checked by a technician.
E007
One or more rear fans failed
RESET according to the instructions.
Check the operation of rear fans and contact Technical Service (see Appendix).
E008
The filling/discharge tank contains water of
inadequate quality.
RESET according to the instructions.
Empty the filling tank and refill it with demineralised / distilled water of adequate
quality (<15 μs/cm).
If an automatic filling system is present, empty the external container and fill it
with water of adequate quality. If a Pure100/500 demineraliser is present, replace
the filter elements.
E009
The filling/discharge tank contains water of very poor
quality.
RESET according to the instructions.
Empty the filling tank IMMEDIATELY and refill it with demineralised / distilled
water of adequate quality (<15
μs/cm). If an automatic filling system is present,
IMMEDIATELY empty the external container and fill it with water of adequate
quality. If a Pure100/500 demineraliser is present, replace the filter elements
IMMEDIATELY.
In these conditions, the steriliser allows starting a maximum of 5
consecutive cycles, after which it locks until the tank is filled with
demineralised /
distilled water of adequate quality (<15 μs/cm) are
replaced. This precaution is necessary to prevent damage to the
device.
E010
Door open (or not properly closed) at program start
(START).
RESET according to the instructions.
Properly close the door and restart the program.
Door position microswitch failure.
Contact Technical Service
(see Appendix).
E011
The discharge tank contains water of very poor
quality.
RESET according to the instructions. Empty both tanks IMMEDIATELY, replace
filter elements and fill the filling tank.
E021
Door lock mechanism limit microswitch failure.
RESET according to the instructions.
Contact Technical Service (see Appendix).
Door lock system gearmotor failure.
E022
Door lock microswitches failure
RESET according to the instructions.
Contact Technical Service (see Appendix).
E030
Water level in the filling tank below minimum.
RESET according to the instructions.
Top up with water up to the MAX. level (or at least up to over the MIN. level).