Preventative Maintenance
Unit testing and validation
The following outlines the recommended regular maintenance to be performed by the user and
technicians for the HYDR
IM
C61w G4 washer.
To ensure the unit's ongoing cleaning efficacy, SciCan recommends observing the following periodic
testing protocols.
The recommended periodic testing protocols are listed in the table below:
When
Who
What Tests
Upon installation
By a SciCan-
approved
technician
Commissioning/Installation testing
as outlined in the installation document
included with the unit.
Daily tests and checks
By the user
•
door lock check
•
wash arm rotation check
•
door seal check
•
load carrier check
•
check and clean chamber filters
•
visual examination (inspection under
magnification) of each load for residual soil.
Annually
By a SciCan-
approved
technician
Annual maintenance schedule as
outlined in the service manual
Annual Service Requirements
The HYDR
IM
C61w G4 is designed to be maintenance free; however, it is recommended that a SciCan-
approved service technician perform an annual check.
The following checks are recommended in order to maintain optimum performance of the unit.
Annual service schedule
• Check integrity of incoming and outgoing services (power, water supply, drain)
• Check general condition of machine
• Replace dryer filter and reset dryer counter (only if required)
• Inspect and replace main chamber
seal (only if required)
• Inspect and replace lower door seal (only if required)
• Check solution container connection for leaks
• Check salt level and replenish as required
• Check water hardness (test strips) and adjust salt regeneration settings if requ
ired.
• Inspect and clean sump filters (check sump for debris)
• Check wash arms for blockages and remove them for cleaning if required.
• Review error history