background image

• Check individual component functionality. Go to the technician menu (enter access code 7919) and 

select ‘Diagnostic Tools’ then select ‘Component Tests’. From here you can scroll through and check the 

functionality of the following components:  

 

• Coolin

g fans 

• Air gap pump

 & valve 

• Chamber heater

 

• Door latch

 

• Salt regeneration valve

 

• Dosing pump

 

• Dryer

 

• Hot water valve

 

• Cold water valve

 

• Waste pump

 

• Check program selection

 

• Check dosing pump volume; dosing pump is pre

-calibrated. Volume cannot

• 

be adjusted.

 

• Reset service cycle counter

 (Maintenance Reset) 

• Clean machine

 

Equipment and parts requirements for annual service

 

• Dryer filter (HEPA) (Part number 01

-

113277S)

 

• Main chamber seal (Part number 01

-

113300S)

 

• Lower door seal (Part number 01

-113661S) 

• Service Manual (Part number 95

-

113023)

 

• Water hardness test strips (Part number 01

-

108305S)

 