LCS Adaptor and Rack
(included)
LCS Plug
(included)
Rack for vertical instruments (optional)
HYDR
IM
®
C61wd G4 Lumens Cleaning System (LCS)
Connecting handpieces to the LCS adaptor
Using the touchscreen menu
Screw the LCS fitting
to the appropriate
handpiece coupling.
Insert the handpiece coupling into the handpiece.
Push the handpiece into one
of the ports.
The HYDR
IM
will
automatically detect
the presence of the
LCS adaptor. Select
the desired program
and follow the prompts.
Upon completion, open
the door and remove
the handpieces from the
adaptor by pushing on the
tab at the side of the port.
Example: Regular cycle
1
2
3
The Lumens Cleaning System
(LCS) Adaptor can only be
used with HYDR
IM
C61wd G4
models equipped with a lumens
cleaning system. If your HYDR
IM
is LCS enabled, you will see a
white plug on the inside back
left of the chamber.
1
2
3
English
Multiflex
Intramatic
Push tab
to release
Setting up the LCS unit
Remove the screw and
pull out the white plug on
the back wall of the unit’s
chamber.
Pull the quick connect
fitting ring back, insert air
line coupling and release
the ring to lock.
Attach the LCS plug to the
HYDR
IM
trolley. (Stays in
the down position when
LCS rack is not being used)
Attach the LCS adaptor and rack
to the trolley, engaging the tab at
the top of the LCS plug. Place the
cannula cleaning basket on the rails
of the LCS rack.
1
2
4
3
Please note: As recommended
by leading manufacturer/suppliers
of handpieces, the hollow parts
of instruments should be dried
immediately after cleaning and
disinfection using medical compressed
air before additional processes or
sterilization. This is to be followed by
recommended regular maintenance
procedures and application of
maintenance agents/oils. Please
follow all applicable health & safety
regulations for your region.