Quick Reference Guide

1

2

5

or

der numb

er 01

-108367S

Caution:

1.

Hydrim Cleaning Solution is compatible with solid 
stainless steel instruments that are resistant to 
heat and alkalinity.

2.

Carbon steel, aluminum, nickel and rhodium-plated 
instruments may corrode in Hydrim Cleaning Solution. 
Do not process burrs or endofiles in the Hydrim.

3.

Do not process stainless steel and carbon steel 
instruments at the same time.

For best results:

Use Stat-Dri Plus rinse aid (part number 8OZPLUS) to prevent spotting.

P0 should be used to keep instruments moist when they will not be washed within one hour.

P1 and P2 should be used to wash instruments in baskets.

P3 and P4 should be used to wash instruments in cassettes and heavily soiled instruments.

Replacing the Hydrim Cleaning Solution:

When the chemical indicator light shines continuously, and Ed appears on the display, the solution level
is low and the container is in need of replacement.

1.

Open the cleaning solution door.

2.

Put a new cleaning solution box into the drawer upside down.

3.

Tear open the tab.

4a.

Pull out the spout and remove the protective cap.

4b.

Attach the hose tightly to the connector on the cleaning solution box.

5.

Place the box in the cleaning solution drawer and then close.

3

Shines when cleaning solution needs changing

Flashes when rinse-aid needs filling

8 min.*

20 min.*

30 min.*

30 min.*

40 min.*

Shines when salt needs refilling

*Actual cycle time will vary depending on 

incoming water temperature and pressure

4a

4b