10
Water supply
High quality distilled water is recommended for use in the STATCLAVE G4. Deionized, demineralized or
specially filtered water can also be used. Never use tap water. The STATCLAVE G4’s water conductivity sensor
automatically reads the water quality and will not allow the unit to run a cycle unless the water quality is
≤
6.4ppm / 10
µ
S (conductivity at 20
˚
C/ 68
˚
F).
Safety information
THE FOLLOWING TERMS APPEAR IN THIS SERVICE MANUAL
CAUTION!
A potential hazard to the operator or to patients.
IMPORTANT!
A situation that may affect the functioning of the unit or lead to a mechanical failure.
TIP
Additional information that may be helpful.
Pay close attention to the following symbols that appear on the unit:
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...