32
Test cycles
Running a vacuum test
The vacuum test checks the sterilizer’s plumbing system for leaks. Run this test with the rack and empty
trays in the chamber.
IMPORTANT: Vacuum tests must be conducted when the unit is cool. Running a vacuum test on a hot
chamber may cause the test to fail.
Running a vacuum test can take a minimum of 45 minutes. When the test is complete, the screen will
display a CYCLE COMPLETE message. If the test has failed, see Troubleshooting.
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...