43
Using the service menu
To access the service menu, select the SETTINGS icon from the home screen, press on the image of the
technician and enter the service code 7919 on the keypad.
Menu item
Submenu item
What to do with it
PRODUCTION TOOLS
Calibration
To calibrate the unit using a
probe, set calibration to ON so
that you can run a cycle and
adjust the temperature offset.
Burn-in mode
Conductivity Setup
DIAGNOSTIC TOOLS
This menu contains a number of tools technicians can use to more accurately diagnose problems.
Device Status/Test
Repeater Mode
Use to run a series of cycles in
succession. Can be set to repeat
1 to 5 cycles in a row or set it to
ON to run the same cycle over
again until you stop it.
Repeater count
Debug Mode
To display the debug screen
while running a cycle, set Debug
Screen to ON.
Clear CF Printouts
Calibration Offsets
To calibrate the unit using a
probe, set calibration to ON so
that you can run a cycle and
adjust the temperature offset.
FACTORY DEFAULT
Cycle Number
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...