MONTHLY or every 100 cycles
Drain and clean both water reservoirs. Remove
and clean both reservoir filters.
Inspect and clean the 3 chamber filters
Clean the external distilled water tank - if installed
Clean the chamber rack and trays
EVERY 6 MONTHS or 500 cycles
(Message appears at intervals)
Perform all the cleaning tasks listed in the monthly
schedule above.
Replace the bacteriological filter
Replace the door seal
TECHNICIAN
EVERY YEAR or 1,000 CYCLES
A complete maintenance of the autoclave
including testing of the pressure relief valve and
the power failure pressure/vacuum relief valve (by
a SciCan-approved technician) is recommended.